Model 2400 Patch Clamp Amplifier
Specifications
Probe Gain and Bandwidth
Feedback Resistor Probe Gain Maximum Current Minimum Bandwidth
10MΩ 10mV/nA 1000nA 100kHz
100MΩ 100mV/nA 100nA 100kHz
1GΩ 1mV/pA 10nA 60kHz
10GΩ 10mV/pA 1nA 40kHz
 
Probe Case Size

2.1" (54mm) long x 0.8"(20mm) high x1.4"(36mm) wide

BNC input connector

1mm ground pin connector

 
Maximum Instrument Noise

Voltage clamp modes measured with the internal 4 pole Bessel filter and minimum external noise.

  Headstage Resistor (Ωs)
Bandwidth 10M 100M 1G 10G
1Khz Theoretical Limit 0.002nA 0.4 pA 0.1 pA 0.04 pA
1kHz 0.003nA 0.5 pA 0.17 pA 0.08 pA

Current clamp modes have noise levels less than 20µV with inputs shorted.

 

Internal Bandwidth

For voltage clamping only the bandwidth is settable from 10kHz to 100kHz.

 

Membrane Current Output Range
Probe Feedback Resistor Output Range
10MΩ 10mV/nA-1V/nA
100MΩ 100mV/nA-10V/nA
1GΩ 1mV/pA-100nV/pA
10GΩ 10mV/pA-1V/pA

 

Membrane Voltage Output Range

10mV/mV-1V/mV

 

Filter

4 pole Bessel filter with 6 pre-set values, and bypass: 0.5 kHz, 1.0 kHz, 2.0 kHz, 5.0 kHz, 10.0 kHz, and 20.0 kHz.

0.1kHz resolution

 

Capacity Compensation

Voltage clamp:

            FAST1: 0-10pF, 0.2-2µs            FAST2: 0-1pF, 0.1-10ms

Current clamp

            FAST1: 0-10pF            No FAST2

 

Whole Cell Compensation

Raccess: 0-100MΩ          Cmembrane: 0-100pF

 
Series Resistance

RsPre: 0-100%

RsComp: 0-100%

LagCoarse: 1-100µs

LagFine: 1-10µs

 

DC Balance

Up to the value of the low feedback resistor in the probe

 

Hold Command

Voltage clamp: ±200mV 

Current clamp:

Probe Resistor 10M 100M
Command Range ±200nA ±20nA

 

Offset Command

Voltage clamp: ±100mV          Current clamp: ±1V

 

Speed Test Command

Used in voltage clamp only.

Probe High Gain:  0.2nA p-p at 60Hz

Probe Low Gain:  1nA p-p at 60Hz

 

Vcomp Command

±5mV at 60Hz

 

Iresist Command

±1nA at 60Hz

 
External Command 

Voltage clamp:

100mV/V with the external switch set to 410, 20mV/V with the external switch set to 450

Current clamp:

Probe Resistor 10M 100M
410 range ±100nA/V ±10nA/V
450 range ±20nA/V ±2nA/V

Catalog
Number
DescriptionPriceQuantity
Ordered
Item
Sub-Total
1 - 45+
880000Model 2400 Patch Clamp Amplifier$2,595.00$2,465.25$0.00
880210Patch Clamp Headstage, 1G/10M Feedback Resistors$305.00$289.75$0.00
880218Patch Clamp Headstage, 10G/10M Feedback Resistors$305.00$289.75$0.00
880222Patch Clamp Headstage, 10G/100M Feedback Resistors$305.00$289.75$0.00
880260Model Cell$95.00$90.25$0.00
Prices generated at 4:26 AM on 11/22/2008.
   
ISO 13485:2003 Certified!

Have you considered?
Soldering Station
Multimeter
BNC Connector

SHOPPING CART

YOUR
ACCOUNT
· Mailing List
· Payment Options
SHIPPING &
SERVICE